The controller can be configured to measure pH or ORP during the installation.
It provides manual and automatic temperature compensation, if a Pt100 is connected.
When the RTD is connected, the instruments provides the temperature reading.
Specifications
- pH range: 0/14.00 pH, accuracy 0.01 pH
- ORP range: -1000/+1000 mV, accuracy 1 mV
- Temperature range: -10/+110 °C, accuracy 0.1 °C
- Analog output: 0-20 or 4-20 mA isolated. Programmable within the scale
- 2 set-point with min/max function, hysteresys and delay time programmable
- Min/max and set points timing alarm relay
- Dual software filter for more steady measures
- 3 software access levels, keyboard lock and access code number
- Input from antimony sensor
- Automatic recognition of the buffer solutions
- Operating mode: automatic and manual
- Temperature readout
- Continuous or flashing alarm relay
- Power supply: 110/220 volt, 5 VA
- Dimensions: 96x96x155 mm. DIN 43700